Persons using assistive technology might not be able to fully access information in this file. For assistance, please send e-mail to: mmwrq@cdc.gov. Type 508 Accommodation and the title of the report in the subject line of e-mail.

Notice to Readers: Medical Equipment Malfunctions Associated with Inappropriate Use of Cleaning and Disinfecting Liquids --- United States, 2007

On October 31, 2007, the Food and Drug Administration (FDA), in collaboration with CDC, the Environmental Protection Agency, and the Occupational Safety and Health Administration, issued a public health notification alerting health-care providers and the public about medical device malfunctions caused by improper use of cleaning and disinfecting liquids.* Inappropriate use of cleaning and disinfecting liquids on certain electronic medical equipment can cause equipment damage and malfunctions, which might have serious, even life-threatening consequences. Under the Safe Medical Device Act, health-care facilities are required to report to FDA any medical device malfunctions that cause or could cause death or serious injury. This notice provides recommendations to help prevent medical device malfunctions attributed to improper cleaning and disinfection.

Cleaning and disinfection are important practices to ensure that medical equipment surfaces do not serve as reservoirs for infectious pathogens. Cleaning is designed to remove infectious pathogens from inanimate objects, whereas disinfection is the process by which remaining pathogens are inactivated. Each of these two distinct processes usually involves the use of liquids (i.e., water and detergents for cleaning and chemical disinfectants for microbial inactivation). Because many types of equipment used in health-care settings have mated surfaces, moving parts, gaps, joints, and unsealed housings, improper cleaning and disinfection can create opportunities for fluids to enter the internal surface of medical equipment, resulting in damage that can cause or contribute to equipment malfunctions.

Health-care facilities, public health officials, and device manufacturers can take several measures to help improve device cleaning and disinfection and to prevent equipment malfunctions in the future. Facility staff should review equipment currently in use to determine which pieces of equipment have manufacturer instructions for cleaning but not for disinfection. Equipment that cannot be disinfected should be used in a way that minimizes the risk for contamination, for example, by positioning it far from contaminated areas or by covering it with a barrier that can be easily cleaned or replaced. If this is not possible, the facility should contact the manufacturer to discuss options for safe and effective disinfection. If the equipment is fluid-tight, and both cleaning and disinfection instructions are provided by the equipment manufacturer, the recommended cleaning agents and chemical disinfectants should be used and the conditions for their use followed. Finally, personnel responsible for cleaning and disinfection must be given appropriate training.

Reports of medical equipment malfunctions that cause or could cause death or serious injury should be made by using FDA's MedWatch 3500A form, available at https://www.fda.gov/medwatch/getforms.htm. Health-care facilities also are encouraged to report medical devices malfunctions that do not meet the mandatory reporting to MedWatch by telephone (1-800-332-1088); by fax (1-800-332-0178); by mail (MedWatch, Food and Drug Administration, 5600 Fishers Lane, Rockville, MD 20852-9787); or online (https://www.fda.gov/medwatch/report.htm).

* Available at http://www.fda.gov/cdrh/safety/103107-cleaners.html.

Use of trade names and commercial sources is for identification only and does not imply endorsement by the U.S. Department of Health and Human Services.


References to non-CDC sites on the Internet are provided as a service to MMWR readers and do not constitute or imply endorsement of these organizations or their programs by CDC or the U.S. Department of Health and Human Services. CDC is not responsible for the content of pages found at these sites. URL addresses listed in MMWR were current as of the date of publication.

Disclaimer   All MMWR HTML versions of articles are electronic conversions from ASCII text into HTML. This conversion may have resulted in character translation or format errors in the HTML version. Users should not rely on this HTML document, but are referred to the electronic PDF version and/or the original MMWR paper copy for the official text, figures, and tables. An original paper copy of this issue can be obtained from the Superintendent of Documents, U.S. Government Printing Office (GPO), Washington, DC 20402-9371; telephone: (202) 512-1800. Contact GPO for current prices.

**Questions or messages regarding errors in formatting should be addressed to mmwrq@cdc.gov.

Date last reviewed: 2/14/2008

HOME  |  ABOUT MMWR  |  MMWR SEARCH  |  DOWNLOADS  |  RSSCONTACT
POLICY  |  DISCLAIMER  |  ACCESSIBILITY

Safer, Healthier People

Morbidity and Mortality Weekly Report
Centers for Disease Control and Prevention
1600 Clifton Rd, MailStop E-90, Atlanta, GA 30333, U.S.A

USA.GovDHHS

Department of Health
and Human Services