Guideline for Disinfection and Sterilization in Healthcare Facilities (2008)

Tables and Figure

Table 7. Minimum cycle times for steam sterilization cycles

Type of Sterilizer Item Exposure time at 250°F (121°C) Exposure time at 270°F (132°C) Drying time
Gravity displacement Wrapped instruments 30 min 15 min 15-30 min
Gravity displacement Textile packs 30 min 25 min 15 min
Gravity displacement Wrapped utensils 30 min 15 min 15-30 min
Dynamic-air-removal (e.g., prevacuum) Wrapped instruments Not Appliciable 4 min 20-30 min
Dynamic-air-removal (e.g., prevacuum) Textile packs Not Appliciable 4 min 5-20 min
Dynamic-air-removal (e.g., prevacuum) Wrapped utensils Not Appliciable 4 min 20 min
Modified from Association for the Advancement of Medical Instrumentation. 813, 819