Table 8. Examples of flash steam sterilization parameters

Type of Sterilizer Load configuration Temperature Time
Gravity displacement Nonporous items only (i.e., routine metal instruments, no lumens) 132°C (270°F) 3 minutes
Gravity displacement Nonporous and porous items (e.g., rubber or plastic items, items with lumens) sterilized together 132°C (270°F) 10 minutes
Prevacuum Dynamic-air-removal Nonporous items only (i.e., routine metal instruments, no lumens) 132°C (270°F) 3 minutes
Prevacuum, Dynamic-air-removal Nonporous and porous items (e.g., rubber or plastic items, items with lumens) sterilized together 132°C (270°F) 4 minutes
Steam-flush pressure-pulse Nonporous or mixed nonporous/porous items 132°C (270°F)
Manufacturers’ instruction
4 minutes
Modified from Association for the Advancement of Medical Instrumentation. 812, 819